ASTM LIS Analyzer Receiver β MCP Server
Let an AI agent collect and query results from clinical and laboratory analyzers (hematology, clinical chemistry, immunoassay, urinalysis, coagulation) that upload results over the standard ASTM E1381 / E1394 host interface (CLSI LIS1-A / LIS2-A2). The server acts as a small, receive-only LIS endpoint: it acknowledges the analyzer's frames, validates checksums, reassembles messages and lets the agent look results up by sample, patient, test or time.
WARNING
Research and lab-operations use only. This is not a medical device and not a validated laboratory information system. It must not be used to report, release or interpret patient results, for diagnosis, or for any clinical decision-making. Results pass through an AI agent that can misread them; always rely on your laboratory's validated LIS/middleware and release process. Patient data is personal health information: patient name, birth date, address and phone are redacted by default; follow your institution's privacy rules before changing that.
| |
|---|
| Package | labmcp-astm-lis |
| Instruments | Analyzers whose host-interface manual specifies ASTM E1381/E1394 (CLSI LIS1-A/LIS2-A2) result upload, e.g. models from Sysmex, Roche (cobas), Abbott, Beckman Coulter, Mindray and others. Check your model's host-interface manual. |
| Interfaces | RS-232 (serial://), TCP/IP with the analyzer as server (tcp://analyzer:port), TCP/IP with the analyzer as client (--option listen_port=5000) |
| Protocol | ASTM E1381 low-level protocol (ENQ/ACK/NAK/EOT, framed text with modulo-8 frame numbers and checksums) + ASTM E1394 records H, P, O, R, C, Q, L. Implemented from the publicly available Beckman Coulter UniCel DxI / Access 2 LIS Vendor Information, C03112-AF (chapters 2β4); the ASTM/CLSI standards themselves are purchase-only. Receive-only. |
| Status | π§ͺ simulated: tested against a wire-level analyzer simulator and loopback TCP, not yet verified on an analyzer. Report a hardware test |
Try it without hardware
uvx labmcp-astm-lis --simulate --check
The simulated analyzer uploads two messages right after start: a CBC with 5-part differential (with H/L flags and suspect-flag comments, one record long enough to need two frames, and one frame corrupted on the line that is NAKed and retransmitted) and a chemistry panel (high glucose and potassium with a hemolysis comment). All patient data in it is fictitious.
Connect your analyzer
On the analyzer, open the host / LIS / "online" communication settings:
- Protocol: ASTM (may be called ASTM E1381/E1394, LIS1-A/LIS2-A2, CLSI or "ASTM host"). Keep the ASTM low-level protocol (ENQ/STX framing) enabled, also over TCP/IP.
- Results: enable automatic result upload.
- Queries and orders: this server does not answer host queries or download orders. Turn off "host query" / "order request" / bidirectional mode, otherwise the analyzer waits for (and times out on) an answer before running samples.
- Link: pick one of
- RS-232: note the baud rate (commonly 9600), 8 data bits, no parity, 1 stop bit (fixed by E1381), and use a null-modem cable.
uvx labmcp-astm-lis --address /dev/ttyUSB0 --check
uvx labmcp-astm-lis --address serial://COM3?baudrate=19200 --check
- TCP, analyzer is the server (it listens on a port): the LIS connects to it.
uvx labmcp-astm-lis --address tcp://192.168.1.80:5000 --check
- TCP, analyzer is the client (you enter the host IP/port on the analyzer): this server listens. Allow the port through the computer's firewall and restrict senders with
allow_from.
uvx labmcp-astm-lis --option listen_port=5000 --option allow_from=192.168.1.80 --check
- Optional: keep a local record of every message with
--option store_path=~/lis/results.jsonl (the newest max_messages are reloaded when the server restarts; if the file cannot be written, messages are still kept in memory and get_connection_status reports store_write_errors).
The receiver starts as soon as the MCP server starts, so uploads are acknowledged even before the agent calls a tool. Received results stay in memory across reconnect (only clear_results or a server restart without store_path removes them), because the analyzer never re-sends a result it has been sent an ACK for.
Add to your MCP client
Claude Code
claude mcp add analyzer -- uvx labmcp-astm-lis --option listen_port=5000 --option store_path=~/lis/results.jsonl
Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, β¦)
{
"mcpServers": {
"analyzer": {
"command": "uvx",
"args": ["labmcp-astm-lis", "--address", "/dev/ttyUSB0"]
}
}
}
Add --read-only to hide clear_results. For other clients, generate the snippet with uvx labmcp config astm-lis --address /dev/ttyUSB0 --client vscode (also cursor, codex, claude-code).
| Tool | Kind | Description |
|---|
clear_results | π control | Delete all received messages and results from this server's memory (e.g. between runs). |
get_command_log | π read | Return the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done. |
get_connection_info | π read | Report which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first. |
get_connection_status | π read | Report the analyzer link state, the messages and results received, and link-layer error counters. |
get_result_detail | π read | Show every decoded field of one result, with its comments, order and patient context, raw records and message header. |
get_results | π read | Return received results, filtered by sample ID, patient ID, test code, time received or abnormal flag (newest last). |
list_received_messages | π read | List the most recent ASTM messages received (newest last). |
reconnect | π safety | Close and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged). |
get_connection_info, get_command_log and reconnect are built into every LabMCP server.
Safety limits
This server is receive-only and never commands the analyzer (it only answers ACK/NAK), so it has no physical safety limits. Privacy settings:
| Option | Default | Meaning |
|---|
redact_patient_info | true | Masks P-record fields 6 (patient name), 7 (mother's maiden name), 8 (birth date), 11 (address) and 13 (telephone) as REDACTED before anything is stored, logged, written to store_path or returned by a tool |
allow_from | (any) | In listen mode, only accept connections from these IP addresses (;- or space-separated) |
max_messages | 5000 | Messages kept in memory; oldest are dropped first |
Patient IDs, sex, attending physician and free-text comment records are not redacted (IDs are needed to filter results). If your analyzers put identifying text in comments, treat the output accordingly.
Example prompts
- "Have any results come in from the hematology analyzer in the last hour? List the samples."
- "Show me all flagged results for sample S26-000123, with units, reference ranges and the analyzer's comments."
- "Export today's potassium results as a table of sample ID, value and flag."
- "Is the analyzer link healthy? How many frames were NAKed today and why?"
- "Compare the glucose results of samples S26-000124 and S26-000131."
- "Clear the stored results before we start the next validation run."
Notes
- Receive-only. No order download, no host-query responses and no result acknowledgements beyond the link-level ACK. Query (Q) messages from the analyzer are stored and reported (
kind: "query") but not answered. Order download is planned future work.
- Link layer (E1381). Every frame is checked: checksum (modulo-256 sum from the frame number to ETX/ETB, as two hex digits), frame number (first frame 1, then modulo 8; a repeat of the last accepted number is treated as a retransmission and ACKed without duplicating data) and framing. Defective frames get NAK and the analyzer retransmits them. Frames outside an ENQ β¦ EOT session are ignored and counted in
bytes_ignored, which usually means the analyzer is not using the ASTM low-level protocol (raw E1394 without framing is not supported). A session that goes silent for 30 s, or restarts with ENQ, has its incomplete message discarded, as E1381 requires; a message that ends with EOT but no terminator (L) record is kept and flagged complete: false.
- Tolerances. Frames with more than 240 text characters and end frames that pack several records are accepted (and counted); a frame whose CR LF never arrives is NAKed after 10 s. Text is decoded as Latin-1 by default (
--option encoding=...; an unknown encoding name is refused at start-up).
- Records (E1394). Delimiters are taken from each header record (
H|\^& by default). Escape sequences &F& &S& &R& &E& and the vendor form &|& are decoded; other escapes are kept verbatim. The test code is the first non-empty "local code" component of the Universal Test ID (^^^TSH, ^^^^WBC), falling back to the first component. The sample ID is the order's Specimen ID, or the first non-empty component of the Instrument Specimen ID. Values are returned exactly as sent ("<0.1", ">500", "Cancelled"); numeric_value is only set for plain numbers.
- Flags and statuses are passed through as sent. Meanings are only attached for the abnormal flags defined by LIS2-A2 / E1394 (
N L H LL HH < > A U D B W) and the result statuses documented in the vendor specification used (F X); anything else is analyzer-specific, so check your analyzer's host-interface manual.
- Timestamps inside records (test started/completed, header time) are the analyzer's local time;
received_at is when this server stored the message (UTC).
- Listen mode serves one analyzer per port: a new connection replaces the previous one (analyzers often reconnect after a reboot without closing the old socket). Run one server entry (or port) per analyzer. There is no TLS; keep the port on the lab network and use
allow_from.
- Only tested against the simulator and loopback TCP. Differences between vendors' ASTM dialects are common; please report what works (or doesn't) with your analyzer.
Hardware verification