Thorlabs Optical Power Meter β MCP Server
Let an AI agent read laser power, log stability and warm-up drift, set the wavelength correction, averaging and range, and dark-zero the sensor on Thorlabs optical power meters through their documented SCPI command set over USB (USBTMC / VISA).
Try it without hardware
uvx labmcp-thorlabs-pm --simulate --check
uvx labmcp-thorlabs-pm --simulate --option sim_sensor=S302C --check
The simulator is a PM100D with an S120C photodiode (or S302C thermopile) looking at a 1.234 mW, 632.8 nm HeNe laser. Setting the wrong wavelength on the photodiode gives a proportionally wrong reading, as on the real bench.
Connect your power meter
- Driver. Thorlabs meters enumerate as USB Test & Measurement Class (USBTMC) devices, vendor ID
0x1313.
- Windows: the Thorlabs driver (TLPM / Thorlabs VISA) is not visible to PyVISA. Install the free NI-VISA runtime, then switch the meter to the NI-VISA driver with the Thorlabs Driver Switcher (installed with the Thorlabs Optical Power Monitor software), and start the server with
?backend=@ivi (see below). This is the route Thorlabs documents for PyVISA.
- Linux: pyvisa-py talks USBTMC directly (installed with this package). Give your user access to the device, e.g.
/etc/udev/rules.d/99-thorlabs-pm.rules:
SUBSYSTEM=="usb", ATTR{idVendor}=="1313", MODE="0666"
then sudo udevadm control --reload-rules and re-plug the meter. NI-VISA also works (?backend=@ivi).
- macOS: pyvisa-py + libusb (
brew install libusb); NI-VISA is an alternative.
- Find the resource string:
uvx labmcp ports (or python -m pyvisa info). It looks like USB0::0x1313::0x8078::P0012345::INSTR (the product ID depends on the model: PM100D 0x8078, PM100USB 0x8072).
- Test the connection:
uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR --check
uvx labmcp-thorlabs-pm --address "visa://USB0::0x1313::0x8078::P0012345::INSTR?backend=@ivi" --check
For the second input of a PM5020 add --option channel=2.
Add to your MCP client
Claude Code
claude mcp add power-meter -- uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR
Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, β¦)
{
"mcpServers": {
"power-meter": {
"command": "uvx",
"args": ["labmcp-thorlabs-pm", "--address", "USB0::0x1313::0x8078::P0012345::INSTR"]
}
}
}
Add --read-only to allow readings but block wavelength, averaging, range and zero changes. For other clients, generate the snippet with uvx labmcp config thorlabs-pm --address <resource> --client vscode (also cursor, codex, claude-code).
| Tool | Kind | Description |
|---|
get_command_log | π read | Return the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done. |
get_connection_info | π read | Report which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first. |
get_device_info | π read | Identify the meter console and the attached sensor head (model, serial, type, wavelength range, power ranges), and report the present wavelength, averaging, range and zero value. |
log_power_series | π read | Log a series of power readings to characterise laser stability, warm-up or drift. Returns mean, stdev, min/max, RMS and peak-to-peak stability (%) and drift (%/min), plus the readings (downsampled to max_points; use save_path, a new .csv file, to keep all of them). |
read_power | π read | Read the optical power once (in W, plus dBm and a formatted string). The reading uses the configured correction wavelength and averaging; set the wavelength first. |
read_sensor_temperature | π read | Read the sensor head temperature in Β°C (thermal sensors and other heads with a built-in temperature sensor only). |
reconnect | π safety | Close and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged). |
set_averaging | π control | Set how many samples the meter averages for each reading. More averaging lowers noise but slows each reading (PM100 series: ~3 ms per sample; PM400/PM101/PM5020: 1 ms). |
set_range | π control | Select auto-ranging, or a fixed power range that fits range_w (the meter picks the most sensitive range that can hold that power). A fixed range reports over-range if exceeded. |
set_wavelength | π control | Set the correction wavelength (nm) the meter uses to convert sensor signal to power. Must be within the sensor's calibrated range (see get_device_info); sent as a whole number of nm. |
zero_sensor | π control | Dark-zero the sensor (removes dark current / thermal offset). BEFORE calling, ask the user to block the beam and cover the sensor aperture completely, and wait for them to confirm: zeroing with light on the sensor corrupts all later readings. Takes a few seconds. |
get_connection_info, get_command_log and reconnect are built into every LabMCP server.
A power meter only measures, so this server has no hazard tools. Your laser is the hazard: wear the right eyewear and never exceed the sensor's rated power or power density.
Safety limits
| Limit | Default | Meaning |
|---|
max_series_duration_s | 600 s | Longest power-logging series an agent may start, estimated as count Γ max(interval, reading time), where a reading takes averaging Γ 3 ms plus ~10 ms |
Override at launch: --limit max_series_duration_s=3600. One call can log at most 3640 s (the tool's timeout minus a margin): longer estimates are refused, and a series that runs slower than estimated stops early and returns what it measured (completed: false). save_path must end in .csv and must not exist yet (an existing file is never overwritten); it is checked before logging starts.
Example prompts
- "Set the meter to 1064 nm and tell me the power after the objective."
- "Log the laser power every 2 s for 10 minutes and report the RMS and peak-to-peak stability. Save the data to ~/data/laser_warmup.csv."
- "I've capped the sensor. Zero it, then I'll remove the cap and you read the power."
- "Measure the transmission of this ND filter: read the power without it, then again after I insert it, and give me the optical density."
- "Is the thermal sensor head still warming up? Read its temperature every minute until it's stable."
Notes
- Wavelength matters. Photodiode responsivity varies by a factor of several across the sensor's range, so a wrong wavelength setting gives a wrong power. The server sends whole nanometres (the PM400/PM101/PM5020 references specify an integer), so 632.8 nm is set as 633 nm. That changes the responsivity by far less than the sensor's calibration uncertainty.
- Units.
MEAS:POW? answers in the meter's configured unit (W or dBm). The driver reads SENS:POW:UNIT? first and always returns watts, so a front-panel unit change can't mislead the agent.
- Out of range. The meter answers INFINITY / 9.9E37 when the signal exceeds a manually selected range. The server reports this as an error that suggests auto-ranging.
- Zeroing (
SENS:CORR:COLL:ZERO) runs in the background. The server polls β¦:ZERO:STAT?, checks the error queue, and returns the zero value (β¦:ZERO:MAGN?). Zero values are lost when the meter reboots.
- Energy (pyroelectric) sensors are recognised but not supported: single-pulse energy measurement needs a trigger/fetch workflow that this server doesn't implement yet.
- Averaging is
SENS:AVER:COUN. One sample takes ~3 ms on the PM100 series and 1 ms on the PM400/PM101/PM5020. The reply timeout scales with the averaging count.
- PM101 and PM5020: Thorlabs' command references for these list exactly the commands used here. The PM5020 adds a channel suffix (
SENS2, MEAS2, SYST:SENS2:IDN?), selected with --option channel=2. Neither has been tested yet, and PM5020 LAN/serial connections have not been tried. The PM100D's Windows TLPM DLL driver is not used; this server speaks SCPI only.
- The simulator's error replies for out-of-range values (
-222,"Data out of range") use the standard SCPI code. The Thorlabs documents only list -113,"Undefined header" explicitly.
Hardware verification